Applications: Soil and Water Quality Assessment, Pathological Analysis, Biomarker Detection, Pesticide Residues Detection
Destructive and non-destructive Organophosphorus pesticide detection from fruits and vegetables.
Successful testing with real samples from Mosambi and Cucumber validated the approach,
Detection Range:- MRL as prescribed in FAO's CODEX guidelines
A low-cost, concise, disposable, miniaturized, and lightweight from the existing expensive equipment’s highly suitable for the low power applications.
The device is successfully tested for the dual detection of ammonia and urea from the blood serum sample.
Detection Range:- For Ammonia 0.5-10 mg L-1 and for Urea: 100-500 mg L-1
Limit of Detection (LoD): For Ammonia: 0.1879 mg L-1 and for Urea: 21.4091 mg L-1.
A Standalone, Point-of-Source, Portable, Low-Cost, Automated & Integrated Microfluidic System for Detection & Quantification of Nitrite
Employs controlled Griess reaction based assay
Low-cost 3D printed syringe pump
Photometric detection employed using LED & photodiode
LOD: 0.07103 ppm; LOQ: 0.21524 ppm (WHO limits: 3 ppm)
Integrated Nitrite, pH and Electrical Conductivity Detection
Uses a multiplexed Polydimethylsiloxane (PDMS) device
Conductivity cell with copper electrodes
Smartphone application for system control
Geo-tagged Data